SLNH-230 Cold Resistance Tester

Overview.

SLNH-230 cold coefficient tester can be used to determine the rubber or film specimens at room temperature and frozen by a certain load, according to the specimen extension length changes and calculate the cold coefficient, or with the required load to the specimen stretched to the same length and calculate the length of the increase in the value of the compression cold coefficient determination. Various technical indicators in line with HG/T3867-2008 or GB/T6035-1985 vulcanized rubber tensile cold coefficient of determination, "HG/T3866-2008 vulcanized rubber compression cold coefficient of determination of chemical standards. SLNH-230 cold coefficient of determination of the use of refrigerant compression cooling, change the previous dry ice and liquid nitrogen cooling method. and liquid nitrogen for cooling, the container with alcohol, so that the surrounding temperature is more uniform, falling temperature faster, saving time, reducing the test energy consumption, reduce test costs.

Technical Parameters.

Test temperature control range

Room temperature to -70°C; Measurement accuracy: ±1°C
Compression test specimen size is round

Diameter Φ10mm, height 10mm±0.2mm

Measurement range

0.01mm-20.00mm; Accuracy: ±0.01mm. computerized display
Tensile specimen size

42mm long X 10mm wide

Timer control range

0-99min59s,Accuracy:±1s
Maximum power consumption

0.6KW

power supply

AC220V±10%,50HZ±1%. single-phase three-wire system, the entire equipment must be well grounded, grounding resistance: ≤2.5Ω
operating environment

Ambient temperature: -10℃ to 30℃.

                 8.2, Altitude: ≤1500m.

                 8.3, Ambient humidity: ≤90%, no condensation.

                 8.4, the presence of sparks and fire ignition in the room is strictly prohibited, requiring fire protection 45 °.

Overall dimensions

(L)1820mmX(W)620mmX(H)1605mm
Tester weight

600Kg